Яндекс.Метрика

A.S. Kudryavtsev, A.L. Makas, M.L. Troshkov,M.L.Grachev, S.P. Podyachev

Том: Talanta
Том: 123 , Год издания: 2014
Многотомное издание: Talanta
Страницы: 140-145

Аннотация

A method for fast simultaneous on-site determination of methyl mercaptan and dimethyl sulfide in air was developed. The target compounds were actively collected on silica gel, followed by direct flash thermal desorption, fast separation on a short chromatographic column and detection by means of mass spectrometer with atmospheric pressure chemical ionization. During the sampling of ambient air, water vapor was removed with a Nafion selective membrane. A compact mass spectrometer prototype, which was designed earlier at Trofimuk Institute of Petroleum Geology and Geophysics, was used. The minimization of gas load of the atmospheric pressure ion source allowed reducing the power requirements and size of the vacuum system and increasing its ruggedness. The measurement cycle is about 3 min. Detection limits in a 0.6 L sample are 1 ppb for methyl mercaptan and 0.2 ppb for dimethyl sulfide
индекс в базе ИАЦ: 041414